Half Autoclavable Micropipette – M

  1. Light weight, ergonomic design.
  2. Digital display clearly reads volume setting.
  3. Wide measurement rang, 0.1ul to 10000ul.
  4. With new material, it can be fully autoclavable.
  5. Ceramic piston, high corrosion resistance.
  6. PTFE sealing ring, more convenient to use.

 

M Series Variable Volume Micropipette Technical Data

Model

Volume Range

Testing

volume

(ul)

Inaccuracy

Imprecision

Increment

(ul)

Tips

Volume

M2.5

0.1-2.5ul

1.25

±3.00%

±0.30%

0.05

10

M10

0.5-10ul

5

±1.50%

±1.50%

0.1

10

M20

2-20ul

10

±1.20%

±1.20%

0.5

200

M50

5-50ul

25

±0.90%

±0.30%

0.5

200

M100

10-100ul

50

±1.00%

±0.15%

1

200

M200

10-100ul

100

±0.70%

±0.15%

1

200

M1000

10-100ul

500

±0.80%

±0.20%

5

1,000

M5000

1000-5000ul

2500

±0.60%

±0.15%

50

5,000

M10000

2000-10000ul

5 ml

±1.20%

±0.20%

0.1 ml

10 ml


 

Whole Autoclavable Micropipette – P

  1. Light weight, ergonomic design.
  2. Digital display clearly reads volume setting.
  3. Wide measurement rang, 0.1ul to 10000ul.
  4. With new material, it can be fully autoclavable.
  5. Ceramic piston, high corrosion resistance.
  6. PTFE sealing ring, more convenient to use.

 

 

 

P Series Micropipette Autoclavable Technical Data

Model

Volume range

Testing Volume (ul)

Inaccuracy

Imprecision

Increment (ul)

Tip volume

P10

0.5 - 10ul

10

±1.00%

±1.00%

0.1ul

20ul

5

±1.50%

±1.50%

1

±2.50%

±2.50%

P20

2 - 20ul

20

±0.90%

±0.40%

0.5ul

200ul

10

±1.20%

±1.00%

2

±3.00%

±2.00%

P50

5 - 50ul

50

±0.60%

±0.30%

0.5ul

200ul

25

±0.90%

±0.60%

5

±2.00%

±2.00%

P100

10 - 100ul

100

±0.60%

±0.15%

1ul

200ul

50

±0.90%

±0.40%

10

±2.00%

±0.10%

P200

20 - 200ul

200

±0.60%

±0.15%

1ul

200ul

100

±0.80%

±0.30%

20

±3.00%

±1.00%

P1000

100 - 1000ul

1000

±0.60%

±0.20%

5ul

1000ul

500

±0.70%

±0.25%

100

±1.50%

±0.70%

P5000

1000 - 5000ul

5000

±0.50%

±0.15%

50ul

5000ul

2500

±0.60%

±0.30%

1000

±0.70%

±0.30%

P10000

2000 - 10000ul

10ml

±0.60%

±0.20%

0.1ml

10ml

5ml

±1.20%

±0.30%

2ml

±1.50%

±0.40%


Multichannel Pipette

  1. 8 and 12 channel pipettes are available for standard 96-well plate.
  2. Dispensing head rotates for optimum pipetting convenience.
  3. Compound material tip cone assemblies allow easy repair and maintenance.
  4. Can be used with universal style pipette tips.

 

 

 

 

 

Multichannel Micropipette Autoclavable Technical Data

Model

Volume range (ul)

Increment (ul)

Test volume (ul)

Max Permissible Systemic Error (Inaccuracy)

Max Permissible Random Error (Imprecision)

%

(ul)

%

(ul)

M8-10

0.5 - 10 ul

0.1

10

1.50

0.150

1.50

0.150

5

2.50

0.125

2.50

0.125

1

4.00

0.040

4.00

0.040

M8-50

5-50 ul

0.5

50

0.50

0.500

0.50

0.250

25

1.00

0.375

1.00

0.250

5

2.00

0.150

2.00

0.100

M8-300

50-300 ul

5

300

0.25

2.100

0.25

0.750

150

0.50

1.500

0.50

0.750

50

0.80

0.750

0.80

0.400

M12-10

0.5-10 ul

0.1

10

1.50

0.150

1.50

0.150

5

2.50

0.125

2.50

0.125

1

4.00

0.040

4.00

0.040

M12-50

5-50 ul

0.5

50

0.50

0.500

0.50

0.250

25

1.00

0.375

1.00

0.250

5

2.00

0.150

2.00

0.100

M12-300

50-300 ul

5

300

0.25

2.1

0.25

0.750

150

0.50

1.500

0.50

0.750

50

0.50

0.750

0.80

0.400