Half Autoclavable Micropipette – M
- Light weight, ergonomic design.
- Digital display clearly reads volume setting.
- Wide measurement rang, 0.1ul to 10000ul.
- With new material, it can be fully autoclavable.
- Ceramic piston, high corrosion resistance.
- PTFE sealing ring, more convenient to use.
M Series Variable Volume Micropipette Technical Data
Model
Volume Range
Testing
volume
(ul)
Inaccuracy
Imprecision
Increment
(ul)
Tips
Volume
M2.5
0.1-2.5ul
1.25
±3.00%
±0.30%
0.05
10
M10
0.5-10ul
5
±1.50%
±1.50%
0.1
10
M20
2-20ul
10
±1.20%
±1.20%
0.5
200
M50
5-50ul
25
±0.90%
±0.30%
0.5
200
M100
10-100ul
50
±1.00%
±0.15%
1
200
M200
10-100ul
100
±0.70%
±0.15%
1
200
M1000
10-100ul
500
±0.80%
±0.20%
5
1,000
M5000
1000-5000ul
2500
±0.60%
±0.15%
50
5,000
M10000
2000-10000ul
5 ml
±1.20%
±0.20%
0.1 ml
10 ml
Model | Volume Range | Testing volume (ul) | Inaccuracy | Imprecision | Increment (ul) | Tips Volume |
M2.5 | 0.1-2.5ul | 1.25 | ±3.00% | ±0.30% | 0.05 | 10 |
M10 | 0.5-10ul | 5 | ±1.50% | ±1.50% | 0.1 | 10 |
M20 | 2-20ul | 10 | ±1.20% | ±1.20% | 0.5 | 200 |
M50 | 5-50ul | 25 | ±0.90% | ±0.30% | 0.5 | 200 |
M100 | 10-100ul | 50 | ±1.00% | ±0.15% | 1 | 200 |
M200 | 10-100ul | 100 | ±0.70% | ±0.15% | 1 | 200 |
M1000 | 10-100ul | 500 | ±0.80% | ±0.20% | 5 | 1,000 |
M5000 | 1000-5000ul | 2500 | ±0.60% | ±0.15% | 50 | 5,000 |
M10000 | 2000-10000ul | 5 ml | ±1.20% | ±0.20% | 0.1 ml | 10 ml |

Whole Autoclavable Micropipette – P
- Light weight, ergonomic design.
- Digital display clearly reads volume setting.
- Wide measurement rang, 0.1ul to 10000ul.
- With new material, it can be fully autoclavable.
- Ceramic piston, high corrosion resistance.
- PTFE sealing ring, more convenient to use.
P Series Micropipette Autoclavable Technical Data
Model
Volume range
Testing Volume (ul)
Inaccuracy
Imprecision
Increment (ul)
Tip volume
P10
0.5 - 10ul
10
±1.00%
±1.00%
0.1ul
20ul
5
±1.50%
±1.50%
1
±2.50%
±2.50%
P20
2 - 20ul
20
±0.90%
±0.40%
0.5ul
200ul
10
±1.20%
±1.00%
2
±3.00%
±2.00%
P50
5 - 50ul
50
±0.60%
±0.30%
0.5ul
200ul
25
±0.90%
±0.60%
5
±2.00%
±2.00%
P100
10 - 100ul
100
±0.60%
±0.15%
1ul
200ul
50
±0.90%
±0.40%
10
±2.00%
±0.10%
P200
20 - 200ul
200
±0.60%
±0.15%
1ul
200ul
100
±0.80%
±0.30%
20
±3.00%
±1.00%
P1000
100 - 1000ul
1000
±0.60%
±0.20%
5ul
1000ul
500
±0.70%
±0.25%
100
±1.50%
±0.70%
P5000
1000 - 5000ul
5000
±0.50%
±0.15%
50ul
5000ul
2500
±0.60%
±0.30%
1000
±0.70%
±0.30%
P10000
2000 - 10000ul
10ml
±0.60%
±0.20%
0.1ml
10ml
5ml
±1.20%
±0.30%
2ml
±1.50%
±0.40%
Model | Volume range | Testing Volume (ul) | Inaccuracy | Imprecision | Increment (ul) | Tip volume |
P10 | 0.5 - 10ul | 10 | ±1.00% | ±1.00% | 0.1ul | 20ul |
5 | ±1.50% | ±1.50% | ||||
1 | ±2.50% | ±2.50% | ||||
P20 | 2 - 20ul | 20 | ±0.90% | ±0.40% | 0.5ul | 200ul |
10 | ±1.20% | ±1.00% | ||||
2 | ±3.00% | ±2.00% | ||||
P50 | 5 - 50ul | 50 | ±0.60% | ±0.30% | 0.5ul | 200ul |
25 | ±0.90% | ±0.60% | ||||
5 | ±2.00% | ±2.00% | ||||
P100 | 10 - 100ul | 100 | ±0.60% | ±0.15% | 1ul | 200ul |
50 | ±0.90% | ±0.40% | ||||
10 | ±2.00% | ±0.10% | ||||
P200 | 20 - 200ul | 200 | ±0.60% | ±0.15% | 1ul | 200ul |
100 | ±0.80% | ±0.30% | ||||
20 | ±3.00% | ±1.00% | ||||
P1000 | 100 - 1000ul | 1000 | ±0.60% | ±0.20% | 5ul | 1000ul |
500 | ±0.70% | ±0.25% | ||||
100 | ±1.50% | ±0.70% | ||||
P5000 | 1000 - 5000ul | 5000 | ±0.50% | ±0.15% | 50ul | 5000ul |
2500 | ±0.60% | ±0.30% | ||||
1000 | ±0.70% | ±0.30% | ||||
P10000 | 2000 - 10000ul | 10ml | ±0.60% | ±0.20% | 0.1ml | 10ml |
5ml | ±1.20% | ±0.30% | ||||
2ml | ±1.50% | ±0.40% |

Multichannel Pipette
- 8 and 12 channel pipettes are available for standard 96-well plate.
- Dispensing head rotates for optimum pipetting convenience.
- Compound material tip cone assemblies allow easy repair and maintenance.
- Can be used with universal style pipette tips.
Multichannel Micropipette Autoclavable Technical Data
Model
Volume range (ul)
Increment (ul)
Test volume (ul)
Max Permissible Systemic Error (Inaccuracy)
Max Permissible Random Error (Imprecision)
%
(ul)
%
(ul)
M8-10
0.5 - 10 ul
0.1
10
1.50
0.150
1.50
0.150
5
2.50
0.125
2.50
0.125
1
4.00
0.040
4.00
0.040
M8-50
5-50 ul
0.5
50
0.50
0.500
0.50
0.250
25
1.00
0.375
1.00
0.250
5
2.00
0.150
2.00
0.100
M8-300
50-300 ul
5
300
0.25
2.100
0.25
0.750
150
0.50
1.500
0.50
0.750
50
0.80
0.750
0.80
0.400
M12-10
0.5-10 ul
0.1
10
1.50
0.150
1.50
0.150
5
2.50
0.125
2.50
0.125
1
4.00
0.040
4.00
0.040
M12-50
5-50 ul
0.5
50
0.50
0.500
0.50
0.250
25
1.00
0.375
1.00
0.250
5
2.00
0.150
2.00
0.100
M12-300
50-300 ul
5
300
0.25
2.1
0.25
0.750
150
0.50
1.500
0.50
0.750
50
0.50
0.750
0.80
0.400
Model | Volume range (ul) | Increment (ul) | Test volume (ul) | Max Permissible Systemic Error (Inaccuracy) | Max Permissible Random Error (Imprecision) | ||
% | (ul) | % | (ul) | ||||
M8-10 | 0.5 - 10 ul | 0.1 | 10 | 1.50 | 0.150 | 1.50 | 0.150 |
5 | 2.50 | 0.125 | 2.50 | 0.125 | |||
1 | 4.00 | 0.040 | 4.00 | 0.040 | |||
M8-50 | 5-50 ul | 0.5 | 50 | 0.50 | 0.500 | 0.50 | 0.250 |
25 | 1.00 | 0.375 | 1.00 | 0.250 | |||
5 | 2.00 | 0.150 | 2.00 | 0.100 | |||
M8-300 | 50-300 ul | 5 | 300 | 0.25 | 2.100 | 0.25 | 0.750 |
150 | 0.50 | 1.500 | 0.50 | 0.750 | |||
50 | 0.80 | 0.750 | 0.80 | 0.400 | |||
M12-10 | 0.5-10 ul | 0.1 | 10 | 1.50 | 0.150 | 1.50 | 0.150 |
5 | 2.50 | 0.125 | 2.50 | 0.125 | |||
1 | 4.00 | 0.040 | 4.00 | 0.040 | |||
M12-50 | 5-50 ul | 0.5 | 50 | 0.50 | 0.500 | 0.50 | 0.250 |
25 | 1.00 | 0.375 | 1.00 | 0.250 | |||
5 | 2.00 | 0.150 | 2.00 | 0.100 | |||
M12-300 | 50-300 ul | 5 | 300 | 0.25 | 2.1 | 0.25 | 0.750 |
150 | 0.50 | 1.500 | 0.50 | 0.750 | |||
50 | 0.50 | 0.750 | 0.80 | 0.400 | |||